Appendix A - Troubleshooting
Fault message Possible causes Suggested corrective measures
CAN’T MAKE SHARP 1. Monitor out of focus. 1. Adjust the focus.
2. Internal splicer error. 2. Service the splicer.
BAD VIEW ANGLE 1. The V-grooves are dirty. 1. Clean the V-grooves.
2. The V-grooves are out of alignment. 2. Service the splicer.
INTERNAL ERROR Splicer needs service. Service the splicer.
CANNOT ESTIMATE LOSS 1. Bad splice picture. 1. Re-fuse the fibers.
2. Wrong parameters used in 2. Check the program and program
program. parameters, and then re-splice.
3. If it happens very frequently, it 3. Service the splicer.
means the optical system is faulty.
BAD SPLICE Unsuccessful splice. Re-splice the fibers.
XCT OUTSIDE RANGE 1. Bad conditions of electrodes 1. Clean electrodes (see p.68)
2. Light emissions from right and 2. Try splicing other fibres to see if
left fibre differ too much problem persists
3. Splicer needs service 3. Service the splicer
General fault messages (cont.)
Fault message Possible causes Suggested corrective measures
SELECT # NOT PRESSED During the threshold type test, the Redo threshold type test.
COMPUTE THRESHOLD? #-button was not pressed, and thus
no new loss data was saved.
NO THRESHOLD FOUND Insufficient data for calculating Re-do the test, paying special attention to
DO SELECT NEXT TIME a threshold type. taking data for splice loss measurement.
NO THRESHOLD FOUND Insufficient data for calculating Reduce FUSION TIME 2 and re-do test.
TRY REDUCE TIME 2 a threshold type.
NO THRESHOLD FOUND Insufficient data for calculating Increase FUSION CURRENT 2 and re-do
TRY INCREASE CURR. 2 a threshold type. test.
NO THRESHOLD FOUND Insufficient data for calculating Reduce FUSION CURRENT 2 and re-do
TRY REDUCE CURR. 2 a threshold type. test.
NO THRESHOLD FOUND Insufficient differences in index Try splicing with the normal splicing
TRY NORMAL PROCESS profiles to calculate a threshold process.
type.
Mode field matching fault messages
When you have gotten a fault message and made note of
what you should do, you can then return to the on-going
procedure by pressing the ESC button.