E-IM54f P23/37
TD830
ELECTROSURGICAL UNIT
If the EPM circuit has been activated.
Remedy:
(i) Because the EPM alarm cannot be reset by the
User, if unit EPM alarm activates as soon as output
is activated, Eschmann Equipment should be
contacted.
If the PVM circuit has been activated.
Remedy:
(i) If unit PVM alarm activates check for accidental
direct contact between the active electrode and an
earthed object.
(ii) Check that the plate electrode is correctly attached
to the patient’s body.
Note: An activated PVM alarm will automatically
be reset by releasing a depressed footswitch or
fingerswitch.
If none of the above remedies stop the alarm activated, or
the (!) alarm is activated contact Eschmann Equipment
and do not use the unit.