screw, which holds the electric outlet plate cover to the socket. DO NOT
OPERATE UNIT WITHOUT PROPER GROUNDING. A fuse, 15AMP, 120VAC, is
located on the power module.
5. CONTROL FUNCTION
A. Main Switch: When unit is ready for operation, put the main
switch to ON and the indicator light on the power and the Control
Module will be ON.
B. Run-Stop Switch: Located on Control Module, must be in the RUN
position for the programmable controller to operate. If during the
program cycle you want to stop the process, turn switch to STOP position
and the program cycle will discontinue. Move switch back to RUN
position and program will START over at the beginning of the fill cycle
when start button is pushed.
Use run-stop switch only for emergency stopping of unit.
C. Fill/Clean/Full Cycle Switch:
Fill: When the Fill cycle is selected and the START button is
pushed the tank will fill and hold with heated detergent. Use this cycle if
you do not have a full load of instruments and you would like to soak
them until ready to complete the cycle.
Clean: Select Clean cycle only after the fill step above has been
run. Pressing START when there is a full load of instruments will
complete the clean/rinse/dry cycle. You cannot start the clean cycle if
the fill cycle (above) is not completed first.
Full Cycle: When Full Cycle is selected and the START button is
pushed the unit will fill with heated detergent and continue
automatically with the clean/rinse/dry steps.
This is a safety precaution so the ultrasonics will not operate without any
liquid in the tank.