b. Recommended separation distance
The device is intended for use in a electromagnetic environment in which RF radiation disturbances are
controlled.
The user or installer of the device can help avoid electromagnetic interference by maintaining a minimum
distance, depending on the maximum power of the radio frequency transmission equipment. Portable RF
communications devices (including devices such as antenna cables and external antennas) must not be
used closer than 30 cm (12 inches) to any part of the device, including cables specified by the
manufacturer. Otherwise, the performance of these devices could be affected.
c
. Electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below. It is up to the
customer or the user to verify that the instrument is used in this environment.
Emissions test Compliance
Electromagnetic environment –
Guidelines
Electromagnetic radiation disturbance (Radiated
Emissions) (CISPR 11)
Group 1
The product uses RF energy for
internal functions.
Disruptive voltage at power stations (Conducted
emissions) (CISPR 11)
Class B
The product may be used in all
establishments, including domestic
sites and those connected directly to
the public low-voltage power.
Harmonic current emission (IEC61000-3-2)
Class A
Complies
Voltage variations, voltage fluctuations and flicker
(IEC61000-3-3)
Complies
d
. Magnetic and electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below.
It is up to the customer or the user to verify that the instrument is used in this environment.
Immunity test
Test level IEC 60601 &
Compliance level
Electromagnetic environment –
Guidelines
Electrostatic discharge (ESD) (IEC61000-4-
2)
± 8 kV contact
± 15 kV air
Environment of a professional
healthcare facility.
Electrical fast transients and bursts
(IEC61000-4-4)
± 2 kV for power supply lines
± 1 kV for the signal ports
Shock Waves (IEC61000-4-5)
± 2 kV in differential mode
± 1 kV in current mode
Assigned industrial frequency magnetic field
(IEC61000-4-8)
30 A/m
Voltage dips, short interruptions and
voltage variations (IEC61000-4-11)
0% U
T
for 0.5 cycles
(0°, 45°, 90°, 135°, 180°,
225°, 270° and 315° for 0.5
cycle)
0% U
T
for 1 cycle
70% U
T
For 25 cycles at 50 Hz
For 30 cycles at 60 Hz
Single phase: 0°
Environment of a professional
healthcare facility. If using the
system requires continued
operation during power cuts, it is
recommended that the medical
device be supplied with a separate
power source (UPS, etc.).
Voltage Interruptions (IEC61000-4-11)
0% U
T
for 250 cycles at 50Hz
for 300 cycles at 60Hz
U
T
is the AC mains voltage before applying the test level.
USER MANUAL > CAUTIONS & WARNINGS
Vision-C 600 - Polarized visual chart system > V1- 02-2021 (VRSQSE)
44

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor Instruments Vision-C 600 and is the answer not in the manual?

Essilor Instruments Vision-C 600 Specifications

General IconGeneral
BrandEssilor Instruments
ModelVision-C 600
CategoryMedical Equipment
LanguageEnglish

Related product manuals