® BPM68 ® BPM68
Care and Maintenance
1. Clean the device and arm cu carefully only with a slightly
moistened cloth.
2. Do not immerse the device in water. It may cause damage if water
enters.
3. Do not use any gas, strong detergent, or solvents to clean the device,
including the arm cu.
4. Do not attempt to disassemble, repair or modify the blood pressure
monitor or arm cu.
5. Disconnect the cu and tubing from the monitor prior to storing.
6. Do not use any liquids or water for cleaning on the monitor or cu.
7. Use a soft, dry cloth to clean your monitor.
8. Do not store in direct sunlight, dust or excessive humidity.
9. Avoid extreme temperatures.
10. Remove batteries if the monitor will not be used for an extensive
period of time.
11. Do not drop or expose the device to heavy shock.
CAUTION:
l
No modication of this equipment is allowed.
l
Keep your equipment out of the reach of children, pets and pest.
l
The patient is an intended operator.
l
Risk of explosion if battery is replaced by an incorrect type.
l
Dispose of used batteries according to the instructions.
Technical Specication
l
Measuring range: Blood Pressure: 30~280 mmHg Pulse Rate : 40~199
beats/min
l
Calibration accuracy: Blood Pressure: ± 3 mmHg Pulse rate: ± 4% of
reading
l
Operating environment: 10°C~40°C (50°F~104°F) with relative
humidity up to 85% (non condensing)
l
Storage/ transportation environment: 20°C~+50°C (-4°F~+122°F)
with relative humidity up to 85% (non condensing)
l
Power source: 4 x 1.5 V AAA batteries
l
Weight: Approx. 226g (exclude batteries)
l
Dimensions: Approx. 95 x 130 x 45(W x H x D)
l
Cu circumference (L Size): approx. 22 ~ 42 cm (9” ~ 17”)\
Error Codes
Err Code Meaning Corrective Action
Err 0 No pulse or detect pulses Take o heavy clothes and
not enough. retry again.
Err 1 Leakage in Cu Pressure/ The arm cu is not fastened
Ination too low. properly. Reapply the cu, and take
a measurement again.
Err 2 Pressure fault. Rest a while, relax and retry again.
Err 3 Deation fault. The arm cu is not fastened
properly. Reapply the cu, and take
a measurement again.
Err Memory error. Take o batteries to reboot the
device, then take another
measurement.
Low batteries. Replace all batteries with new ones.
1716