NOTICE
This information applies only to the
MaxTester Series units.
Replacing Batteries
Your unit uses a lithium-polymer battery that has been especially designed
for EXFO. For this reason, you can only replace the main battery with a
battery designed for your MaxTester unit.
You can purchase new batteries from EXFO.
WARNING
Do not throw batteries into fire or water and do not short-circuit
the battery’s electrical contacts. Do not disassemble.
WARNING
To avoid serious injuries as well as irreparable damage to your
unit, ALWAYS TURN OFF YOUR UNIT (shutdown mode, not
suspend) and DISCONNECT IT FROM THE AC POWER.
Be careful not to drop metal objects such as screws inside the
unit. These could cause a short-circuit leading to a fire or an
explosion.
CAUTION
Electrostatic discharge (ESD) damage can cause complete or
intermittent equipment failures.
Always use an ESD-preventive wrist or ankle strap when
handling the unit while it is open. Ensure that the antistatic
strap makes good skin contact and that the end of its wire is
grounded properly.
All the components that you should be handling are identified
in the procedure hereafter. Do not touch any other component
inside the unit either with tools or your fingers.
To avoid damaging the new battery, handle it with care.
CAUTION
Ensure that no foreign object falls inside the unit. This may cause
irreparable damage to your unit.

Other manuals for EXFO MaxTester series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EXFO MaxTester series and is the answer not in the manual?

EXFO MaxTester series Specifications

General IconGeneral
Display7-inch touchscreen
ConnectivityUSB, Ethernet, Wi-Fi
Operating Temperature-10 °C to 50 °C (14 °F to 122 °F)
CategoryOTDR
Wavelengths1310 nm, 1550 nm
Event Dead Zone0.8 m
iOLMYes
ApplicationFiber optic network testing and troubleshooting
Humidity5% to 95% non-condensing

Related product manuals