DISPLAY BACKLIGHT
Press and hold the button to turn on the display backlight function. The backlight will automatically turn off after
15 seconds.
BATTERY CHECK
The CHECK function tests the condition of the 9V battery. Set the function switch to the 200VDC range and
press the CHECK button. If the reading is less than 8.5, battery replacement is recommended.
HOLD
The hold function freezes the reading in the display. Press the HOLD key momentarily to activate or to exit the
hold function.
AUTO POWER OFF
The auto off feature will turn the meter off after 15 minutes.
LOW BATTERY INDICATION
If the icon appears in the display, the battery voltage is low and the battery should be replaced.
WRONG CONNECTION INDICATION
The icon will appear in the upper right conner of the display and the buzzer will sound whenever the positive
test lead is inserted into the 20A or uA/mA input jack and a non-current (green) function is selected. If this occurs,
turn the meter off and reinsert the test lead into the proper input jack for the function selected.
Specifications
Function Range Resolution Accuracy
200mV 0.1mV
(0.3% reading + 2 digits)
2V 0.001V
200V 0.1V
(0.5% reading + 2 digits)
DC Voltage
(V DC)
1000V 1V
(0.8% reading + 2 digits)
50 to 400Hz 400Hz to 1kHz
2V 0.001V
(1.0% reading +6 digits (2.0% reading + 8 digits
200V 0.1V
(1.5% reading +6 digits (2.5% reading +8 digits
AC Voltage
(V AC)
750V 1V
(2.0% reading +6 digits (3.0% reading +8 digits
200A 0.1A
200mA 0.1mA
(1.5% reading + 3 digits)
DC Current
(A DC)
20A 0.01A
(2.5% reading + 3 digits)
50 to 400Hz 400Hz to 1kHz
200mA 0.1mA
(1.8% reading +8 digits (2.5% reading +10 digits)
AC Current
(A AC)
20A 0.01A
(3.0% reading +8 digits) (3.5% reading +10 digits)
200 0.1 (0.8% reading +4 digits)
2000 1 (0.8% reading +2 digits)
20k 0.01k
200k 0.1k
(1.0% reading +2 digits)
Resistance
20M 0.01M (2.0% reading +5 digits)
-20 to 750ºC 1ºC Temperature
-4 to 1382ºF 1ºF
(3.0% reading +3 digits)
(meter only, probe accuracy not included)
NOTE: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.
NOTE: Accuracy is stated at 18
o
C to 28
o
C (65
o
F to 83
o
F) and less than 75% RH.