TB400-en-GB_V1.3 8/16
4
Calibration
Acompletecalibrationwillrequiretestingwithboth0NTUand100NTUtestsolutions.
Ensurethatthesidesofthebottlearecleananddrytopreventmeasurementerrors.
CalibrationProcedure(0NTUand100NTU)
Note:Makesureyoucompletethecalibrationstepswithaslittletimeaspossiblebetween
the0NTUandthe100NTUtestbottles.
1. InserttheZeroNTUsolutionbottleintothetestingchamber,lineupthewhitemarks,and
closeandlatchthecover.TurnonthemeterandPressandholdthe"CAL"buttonuntilthe
LCDdisplays"CAL"thenreleasethe“CAL”button.
2. WaitonesecondandthentheLCDwilldisplay:
Note:IftheLCDdoesnotdisplayzerowhenusing"0NTU"liquid,pressthe"ZERO"button
continuouslyuntiltheLCDdisplays
azerovalue.
3. Themeterisnowreadyfor"0NTU"calibration.
4. Momentarilypress"CAL"button.TheLCDwillflashCALforabout10secondsandthenshow
thefollowingdisplay:
5. TheLCDwillthendisplay:
6. Liftthelidonthetestchamberand
removethe0NTUtestsolutionbottle.
CAL
ntu
0.00
CAL ntu
CAL
0.00 ntu
100
CAL ntu
w ww . . co m