6 TM500-EU-EN v1.1 12/13
SettheTemperatureUnitsofMeasure(°Cor°F)
1. Accessthet‐CFparameter.
2. Usethearrowbuttonstoselect°Cor°F.PressENTERtoconfirmsetting.
3. PresstheENTERbutton toexittotheSetupmode.PressESCtoexitsetupmode.
SettingtheDataloggerSamplingTime(Rate)
1. AccesstheSP‐tparameter.
2. Usethearrowbuttonstoselectthedesiredsamplingrate.Theavailablesettingsare:
0,1,2,
5,10,30,60,120,300,600, 1800,and3600seconds .Use‘0’formanuallogg ingmode.A
samplingrate>1secondisrecommended.
3.
PresstheENTERbuttontoconf irmtheentry.
4. PresstheENTERbutton toexittotheSetupmode.PressESCtoexitsetupmode.
SDCardFORMATTING
1. AccesstheSdFparameter.
2. UsethearrowbuttonstoselectYEStoformatthecard(selectNOtoabort).Notethatall
dataonthecardwillbelostifformattingisattempted.
3. PressENTERtoconfirmselection.
4. PressENTERagaintore‐confirm.
5. PresstheENTERbuttontoexi
ttotheSetupmode.
PressESCtoexitsetupmode.
Note:alwaysformatanewSDmemorycardbeforeuse.