ZerotheParticlecountSensor
Particles must be purged (removed) from the sensor before and after each use in a high
samplecountenvironment.
Zerothesensor:
1. UnscrewandremovetheIsokineticprobeandattachtheZerocountfilter.
2. TurnonthemeterandintheSetupmode,
a. SettheSamplemode
toCumulative.
b. SettheChannelDisplaytoensureallchannelsareselected.
c. SetSampletimeto60seconds
d. SetSampleCycleto10
3. Startthemeter
4. Allowthemetertorununtilalloftheparticlecountsreadzero.Youmayneedto
runthemeter
multipletimestoensurethatallchannelsreadzero.
5. Turnoffthemeter.
6. RemovetheZerocountfilterandattachtheIsokineticprobe.