Error Code Error Buffer Text Description of Error Possible Problem
003 A03: Meter data CRC Invalid calibration data Meter not calibrated or data corrupt
040 A40: Probe read Probe EPROM read error Probe EPROM missing or defective
043 A43: Probe CRC 1 Probe CRC failed Probe EPROM defective
044 A44: Probe hi current Incorrect probe current Probe EPROM error or meter malfunction
045 A45: Probe CRC 2 Probe CRC failed Probe EPROM not programmed
051 A51: Probe status cmd CRC Probe status cmd CRC error Probe EPROM defective
052 A52: Probe status msg CRC Probe status msg CRC error Probe EPROM defective
053 A53: Probe page cmd CRC Probe page cmd CRC error Probe EPROM defective
054 A54: Probe page data CRC Probe page data CRC error Probe EPROM defective
098 A98: Probe data timeout Probe data timeout Internal meter error
099 A99: Cal data timeout Calibration data timeout Internal meter messaging / reset problem
060 D60: A/D ADC error Internal meter problem
061 D61: A/D zero AD zero error Internal meter problem
062 D62: Chip comm PIC-DSP com error Internal meter problem
063 D63: Offset D/A DAC offset error Internal meter problem
064 D64: I hall D/A Probe current error Broken probe or meter problem
065 D65: Output D/A Output DAC error Internal meter problem
066 D66: Gain Amplifier gain error Internal meter problem
067 D67: High input res High input resistance Broken probe or meter problem
Most errors result from a damaged probe. If an error code related to a probe problem appears, substitute another
probe if one is available to help determine if the probe may be defective.
MODEL 5200 ERROR CODES THAT MAY BE DISPLAYED IF THERE IS A PROBLEM