EasyManua.ls Logo

Ferno iNLINE - Maintenance; Maintenance Schedule; Disinfecting the Fastening System; Cleaning the Fastening System

Ferno iNLINE
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
© Ferno-Washington, Inc. / March 2016 / 234-3621-00
15
6 MAINTENANCE
Maintenance
6.1 Maintenance Schedule
The fastening system requires regular maintenance. Set up and follow a
maintenance schedule. The table at right represents minimum intervals
for maintenance.
Keep maintenance records. For a sample record sheet, see “Maintenance
Record” on page 23.
When using maintenance products, follow the manufacturers directions
and read the manufacturers’ material safety data sheets.
6.2 Disinfecting the Fastening System
Wipe all surfaces with disinfectant. Follow the disinfectant
manufacturers instructions for application method and contact time.
Ferno recommends you inspect the fastening system for damage as
you disinfect.
6.3 Cleaning the Fastening System
1. Hand clean all surfaces of the fastening system with warm water
and a mild detergent.
2. Rinse with warm, clear water. Dry the fastening system with a
towel or allow it to air-dry.
6.4 Inspecting the Fastening System
Have your service’s equipment maintenance personnel inspect the
fastening system regularly.
Follow the checklist at right and operate the fastening system through
all its functions as described in this manual. See “Responsibility of User
to Maintain a Safe Fastening System on page 7.
If inspection shows damage or excessive wear, remove the fastening
system from service until repair is made. See “Parts and Service, Accessories
on page 20.
WARNING
Improper maintenance can cause injury. Maintain the
fastening system only as described in this manual.
Important
Disinfectants and cleaners containing bleach, phenolics, or
iodines can cause damage. Do not use products containing
these chemicals.
Using abrasive cleaning compounds or applicators can cause
damage. Do not use abrasive materials to clean the fastening
system.
If the required backing plates are not present, take the fastening
system out of service and have the backing plates installed
immediately. Then, install a new fastening system.
Make sure all connections are tight, but do not overtighten
the hardware. Overtightening, especially with powered tools,
can cause the nut and bolt beneath the ambulance oor to pull
through the backing plate.
If inspection indicates the components are pulling through
the backing plate, take the fastening system out of service
immediately and install a new fastening system with new
backing plates.
Minimum
Maintenance Each Use As Needed Monthly
Disinfecting (this page)
Cleaning (this page)
Inspecting (this page)
Inspection Checklist
● Are all components present?
● Is the fastening system free of excessive wear?
● Are all parts in good condition (no cracks, corrosion, or
damage)?
● Do all moving parts operate smoothly and properly?
● Does the fastening system lock properly to the mounting rail?
● Does the product load and unload properly?
● Does the product lock properly into the fastening system?
● Monitor the ambulance oor, including beneath the
ambulance, and check the integrity of the fastening system.
Is the system worn? Has exposure to the environment
caused noticeable damage?
● Are the required backing plates present, and are they
securely welded to a primary structural member of the
ambulance oor?
● Is all hardware tight? (Reminder: Do not overtighten the
hardware.)

Related product manuals