© Ferno-Washington, Inc. / March 2016 / 234-3621-00
9
3 OVERVIEW
3.1 Description
The Ferno
®
iN∫LINE™ Fastening System is a device designed to secure
a compatible Ferno ambulance cot or iN∫X® transport system inside a
ground-based ambulance. See “Fastening System Compatibility” on
page 4 and “Compatible Ferno Products” below.
Three lengths are available: iN∫LINE-S, iN∫LINE-M, and iN∫LINE-L.
In this manual, the term fastening system is used to refer to all models of
the iN∫LINE. When dierences exist, the specic model name (iN∫LINE-S,
iN∫LINE-M or iN∫LINE-L) is used.
The fastening system is for professional use only by trained Emergency
Medical Service operators. See “Operator Skills and Training” on page 8.
FEATURES
● Aluminum Construction: The exterior is impact-resistant and easy
to clean.
● Automatic Charging with ICS: The charging system is integrated
into the fastening system and charges or maintains the charge
in the iN∫X battery whenever the iN∫X is secured in the fastening
system. The iN∫X power switches to OFF when it is secured in a
powered ICS.
● Automatic Switching (Charging): If the ambulance electrical
system allows automatic switching, the ICS can use the ambulance
electrical system when the ambulance is running, or utility power
when the ambulance is parked and connected to utility power.
● Safety Hook: The integrated safety hook functions as a safety stop
to hold the cot or iN∫X secure inside the ambulance while operators
raise and lower the legs during the loading and unloading process
● Release Button: Single-press button unlocks the cot or iN∫X for
unloading.
● Tested and Certied for Worldwide Standards Compliance: See
"Compliance" on page 6.
● Track Loading: Allows loading of the cot or iN∫X from a variety of
angles.
Overview
WARNING
Improper use can cause injury. Use the fastening system
only for the purpose described in this manual.
An improper cot/transport system can cause injury. Use
only a compatible Ferno product designed or properly
modied for use with the fastening system.
Installation Overview
To install the fastening system, follow the instructions in an
iN∫LINE installation manual. A manual is provided with each
installation kit. For additional manuals, contact Ferno. See
“Ferno Customer Relations” on page 2.
As a user, be aware of the following:
● The ambulance must comply with the standards of the
country in which it is used. See “Compliance” on page
6 for a list of standards met by the fastening system.
● Installation requires the use of backing plates beneath
the ambulance oor at all mounting points. If your
ambulance does not have the required backing plates,
have them installed.
● Inspect the fastening system regularly to ensure it
meets all required standards required by your local
government. See “Inspecting the Fastening System” on
page 15.
● Install a new fastening system when remounting an
ambulance box. Removing and reinstalling an old
fastening system can weaken the fastening system.
● If the ambulance is involved in a trac accident,
inspect it. See “Inspecting the Fastening System” on
page 15. Replace any damaged components.
3.2 Compatible Ferno Products
The fastening system is compatible only with Ferno products. See
“Fastening System Compatibility” on page 4.
● Ferno® iN∫X® Patient Transport & Loading System™