Cardiac stimulator 4427 Easypace - Instructions for use
8/40
• Before each use check the integrity of the device and all its accessories. Particular
attention must be paid when inspecting the cables and connectors. In case of doubt do
not use the device and contact FIAB.
• Before each use check the compatibility of the pin of the electrocatheter with the
socket of the device.
In the event of using electrocatheters not fitted with protected pins, but instead with pins with male contacts with a
diameter of less than 2 mm, it is recommended using the heart surgery cable F7818/30 with screw-on sockets in
order to allow for stimulation.
• Insert the terminals of the electrode fully into each socket without forcing, following the
indications in chapter 10 - "Connection to the patient and Instruction for use”.
• Before each use check that the transparent protective cover of the keypad is correctly
inserted and that there are no breaks or tears. This protection must always be put back
into place after setting the parameters.
• Should the device be dropped, check its correct functioning and in case of doubt
contact FIAB before reusing. If there is any apparent mechanical damage avoid using
the device and contact FIAB.
• All operations to open the device, gain access to the internal parts, and repair
the same must only ever be carried out by authorized FIAB personnel.
• The environment where the device is used must be compliant with the safety
regulations.
• The device must never be immersed in water or any other liquids.
• It will be necessary to remove the batteries from their compartment if the stimulator is
stored for prolonged periods (longer than 4 weeks).
• After storing the 4427 device in environments with temperatures and/or humidity
unsuited to its use, before using again wait until a suitable temperature and humidity
have been reached as indicated on the “characteristics” card.
• Product not sterile and not sterilizable: the hygienic conditions of the device cannot be
guaranteed in the event of being placed inside a sterile field.
• In order to avoid damaging the device it must not be sterilized. To clean the device and
the accessories follow the indications in chapter 12.
• The accessories supplied with the device are not sterile.
IT IS RECOMMENDED HAVING THE 4427 DEVICE CHECKED AT LEAST ONCE A YEAR BY OUR
MEASURING LABORATORY.