7.6 Drying
Manual drying: Removing any liquid residue with a lint-free cotton cloth, and then
blow dry with compressed air (1-2 Bar)
Automatic drying: Refer to chapter 7.5 f)
7.7 Inspection and Maintenance
After cleaning and disinfection, visually inspect the Lip Hook, File Clip and Contra
Angle Holder. If no visible contaminants are found, it means that the Lip Hook, File
Clip and Contra Angle Holder have been cleaned.
If damage is found by visual inspection, stop using it and purchase damaged parts
from the manufacturer or dealer.
7.8 Package
Immediately after drying, put the Lip Hook, File Clip, Contra Angle Holder/Lighting
device and Contra Angle into the steam sterilization bag for sealed packaging.
Caution
Steam sterilization bag should comply with ISO 11607-1 and must be sealed
with a sealing machine.
7.9 Sterilization
Use a high pressure steam sterilizer in accordance with ISO 17665-1 for
sterilization.
a) Sterilization parts: The Lip Hook, File Clip, Contra Angle Holder/Lighting device
and Contra Angle
b) Sterilization method: Autoclave
c) Sterilization conditions: 134 °C for not less than 5 minutes