Cannot charge
the battery
The adapter is not
reliably connected
Check that the adapter
connection is reliable
The battery is
running out
quickly
The charging time for the
battery is too short
Charging time for more than
5 hours
Apex locator
imprecise
/ not sensitive
Test wire connection
unreliable
Reconnect the test wire or
you can contact the file clip
to lip hook directly to check
the connection status
The test wire has an open
circuit or a short circuit
The root canal is in poor
condition
Cannot start the
motor/ motor
does not work
Clean or replace the contra
angle
When the motor
is running, the
torque value is
high
Contra angle wear,
resistance becomes larger
Enter the setup mode and
run the calibration
procedure. If the calibration
fails, replace the contra
angle