The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hy-
pochlorite) specic for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of
potable water and disinfectant, according to the proportions indicated on the
packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation
of air bubbles on the parts. Leave the parts immersed for the amount of time in-
dicated on the packaging of the disinfectant associated with the concentration
chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant
manufacturer.
method B: sanitise the accessories
C1.1-C1.1a-C1.1b-C1.1c-C1.2-C1.3-C1.4-C2-C3-
C4
by boiling them in water for 10 minutes; use demineralised or distilled water to
prevent calcium deposits.
method C: sanitise the accessories
C1.1-C1.1a-C1.1b-C1.1c-C1.2-C1.3-C1.4-C2-C3-
C4
with a hot steam steriliser for baby-bottle (not the microwave type). Perform the
process faithfully following the instructions of the steriliser. To ensure that the disin-
fection is eective, choose a steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them
on a paper towel. Alternatively, dry them with a jet of hot air (for example, from a
hair dryer).
At the end of each use store the device complete with accessories in a dry place
away from dust.
AIR FILTERING
The unit is equipped with an air lter (A3) which should be
replaced when dirty or when its colour changes. Do not wash
or re-use the used lter. The lter must be regularly replaced to
help ensure ecient compressor performance. The lter should
be regularly checked. For spare lters contact your dealer or an
authorised service centre. To replace the lter:
Pull out the dirty lter with common tweezers and push the
new one all the way in its seat. Do not replace the lter during use. Only use original
accessories and spare parts by Flaem, we disclaim any liability in the event of
using non original spare parts or accessories.
7