Ronald CalderonAug 12, 2025
What to do if the Flexicon Laboratory Equipment cap is not crimped tight or is crimped too tight?
- Bblake84Aug 12, 2025
If the cap is not crimped tight or is crimped too tight on your Flexicon Laboratory Equipment, it could be due to several reasons. First, the crimp head height might be incorrect; adjust it slightly, keeping in mind that lower is tighter and higher is looser. Second, the stroke length could be incorrect, requiring an adjustment of the stroke. Finally, dirty or worn jaws can also cause this issue; clean and polish the jaws. If the problem persists, contact your supplier or WMF.