What to do if there is Checksum error of EPL printout format in Flintec Accessories?
- KKaren CarrollSep 12, 2025
If a checksum error of EPL printout format occurs, press enter key and re-load the EPL code, or change E2ROM ( U13 ).
What to do if there is Checksum error of EPL printout format in Flintec Accessories?
If a checksum error of EPL printout format occurs, press enter key and re-load the EPL code, or change E2ROM ( U13 ).
What to do if the weight is too low for Flintec FT-111 Accessories?
If the weight is too low, check the load, as the load cell or instrument could be broken.
What to do if there is a communication error between the display board and main board in Flintec Accessories?
If there is a communication error between the display board and the main board, power off the instrument and reenergize it after 30 seconds. If the issue persists, consider changing the main board.
What to do if there is ADC initialization error in Flintec FT-111 Accessories?
If an ADC initialization error occurs, where the ADC could not initialize at power on due to a malfunction in the ADC or its interface circuitry, power off the instrument and reenergize it after 30 seconds. Also, check the external load cell connection and the load cell connector in the instrument. If problems persist, consider changing the main board or second scale board.
What to do if there is The DLC Board is broken or not installed in Flintec FT-111?
If The DLC Board is broken or not installed, re-energize indicator or change the DLC Board.
What to do if there is E2PROM Error in Flintec FT-111 Accessories?
If there is an E2PROM Error, change main board.
What to do if the weight is out of power on zero range in Flintec FT-111 Accessories?
Press key to start indication without zeroing and call service.
What to do if there is Alibi memory could not be initialized in Flintec FT-111 Accessories?
If the alibi memory could not be initialized, check the alibi memory SD card, or change the main board.
What to do if there is a Checksum error in Flintec FT-111 Accessories?
If a checksum error occurs, check the termination resistors and the DLC connection (RS-485 & Power supply) hardware.
What to do if the load cell output is decreased after loading in Flintec FT-111?
If the load cell output decreases after loading, check the load cell connections and the test weight loading. You may also need to recalibrate the scale or change the mainboard.
Brand | Flintec |
---|---|
Model | FT-111 Series |
Category | Accessories |
Language | English |
Comprehensive guide to various electrical connection options for the instrument.
Covers fulfilling data recording obligations using the alibi memory function.
Overview of the programming menu structure and parameter adjustment.
Detailed steps for performing scale calibration and adjustments.
Process for addressing digital load cells and shift adjustment methods.
Covers Modbus RTU and TCP/IP interfaces, data structures, and setup.
Details Ethernet TCP/IP interface parameters, data formats, and setup.
Information on Profibus DPV1 interface, baud rates, LEDs, and data format.
Details the Profinet interface, LEDs, parameters, and GSDML configuration.
Covers CANopen network communication, LEDs, and data format.
Details EtherNet/IP interface, LEDs, parameters, and EDS configuration.
Covers EtherCAT interface, LEDs, data format, and ESI configuration.
Details CC-Link network, LEDs, configuration, and data format.