What to do if there is Checksum error of EPL printout format in Flintec FT-111 D Accessories?
- MMr. Terry Brown MDSep 12, 2025
If a checksum error of EPL printout format occurs, press enter key and re-load the EPL code, or change E2ROM ( U13 ).
What to do if there is Checksum error of EPL printout format in Flintec FT-111 D Accessories?
If a checksum error of EPL printout format occurs, press enter key and re-load the EPL code, or change E2ROM ( U13 ).
What to do if the weight is too low for Flintec FT-111 D?
If the weight is too low, check the load, as the load cell or instrument could be broken.
What to do if there is a communication error between the display board and main board in Flintec Accessories?
If there is a communication error between the display board and the main board, power off the instrument and reenergize it after 30 seconds. If the issue persists, consider changing the main board.
What to do if The DLC is over in Flintec Accessories?
If The DLC is over, check mechanical installation and DLC.
What to do if there is The DLC Board is broken or not installed in Flintec FT-111 D Accessories?
If The DLC Board is broken or not installed, re-energize indicator or change the DLC Board.
What to do if there is No regular response from load cell in Flintec FT-111 D Accessories?
If there is no regular response from the load cell, reenergize the indicator, check the DLC connection (RS-485 & Power supply), or change the load cell.
What to do if the weight is out of power on zero range in Flintec FT-111 D Accessories?
Press key to start indication without zeroing and call service.
What to do if there is Alibi memory could not be initialized in Flintec Accessories?
If the alibi memory could not be initialized, check the alibi memory SD card, or change the main board.
What to do if there is The address and S/N of the load cell do not match in Flintec FT-111 D?
If the address and S/N of the load cell do not match, check the DLC and S/N, reenergize the indicator, or readdress the DLC.
What to do if the load cell output is decreased after loading in Flintec FT-111 D Accessories?
If the load cell output decreases after loading, check the load cell connections and the test weight loading. You may also need to recalibrate the scale or change the mainboard.
Detailed wiring instructions for load cells and other interfaces.
Wiring details for connecting analogue load cells to the FT-111.
Wiring details for connecting digital load cells to the FT-111 D.
Configuration and pinout for the Ethernet TCP/IP interface.
Configuration and pinout for the EtherNET/IP interface.
Details on activating and using the Alibi memory for data recording.
Navigating the programming menu and modifying parameters.
Procedures for performing scale calibration with test weights.
Steps for addressing and configuring digital load cells.
Details on Modbus RTU and TCP/IP communication protocols.
Configuration parameters for the Ethernet TCP/IP interface.
Information on Profibus DP interface, baud rates, and LEDs.
Details on the Profinet interface, LEDs, and configuration.
Information on CANopen network communication, LEDs, and data format.
Information on EtherCAT interface, LEDs, and data format.
Details on CC-Link network communication, LEDs, and configuration.
| Brand | Flintec |
|---|---|
| Model | FT-111 D |
| Category | Accessories |
| Language | English |