Questions and Answers

  • C
    Christopher TerryAug 23, 2025
    What to do if the load exceeds the operation range in Flintec Accessories?
    • M
      Mary CastroAug 25, 2025
      If the load exceeds the operation range, check the load and the calibration. It is also possible that the load cell or instrument is broken.
  • D
    dyerjessicaSep 1, 2025
    What to do if there is a Checksum error in Flintec FT-111 D?
    • L
      laura72Sep 1, 2025
      If a checksum error occurs, check the termination resistors and the DLC connection (RS-485 & Power supply) hardware.
  • J
    Joyce MayerSep 1, 2025
    What to do if there is Alibi CSUM error in Flintec FT-111 D?
    • B
      Brandon YangSep 1, 2025
      If an Alibi CSUM error occurs, check the alibi memory records. If necessary, change the Alibi memory SD card or the main board.
  • M
    Mr. David WilliamsSep 2, 2025
    What to do if there is ADC output is not changed for the span calibration in Flintec Accessories?
    • D
      David SmithSep 2, 2025
      If the ADC output is not changed for the span calibration, recalibrate the scale or change the mainboard.
  • C
    Christopher HoweSep 3, 2025
    What to do if there is Clock error in Flintec FT-111 D Accessories?
    • R
      Robert OrrSep 3, 2025
      If a clock error occurs, change the CR2032 battery or the main board.
  • R
    Rebecca CameronSep 4, 2025
    What to do if there is Wrong SD card at alibi memory in Flintec FT-111 D Accessories?
    • N
      nadamsSep 4, 2025
      If the wrong SD card is used at alibi memory, order an Alibi SD card.
  • Z
    zscottSep 4, 2025
    What to do if there is Status error of load cell in Flintec Accessories?
    • P
      Pamela ThomasSep 5, 2025
      If there is a status error of the load cell, change the load cell.
  • D
    Douglas WalkerSep 5, 2025
    What to do if the weight is out of power on zero range in Flintec FT-111 D Accessories?
    • J
      Jorge JordanSep 5, 2025
      Press key to start indication without zeroing and call service.
  • S
    Sarah MooreSep 6, 2025
    What to do if there is Alibi memory could not be initialized in Flintec Accessories?
    • R
      Richard TravisSep 6, 2025
      If the alibi memory could not be initialized, check the alibi memory SD card, or change the main board.
  • K
    Kari HowardSep 7, 2025
    What to do if Alibi SD card is not supplied from FLINTEC in Flintec FT-111 D Accessories?
    • J
      James WilliamsSep 7, 2025
      If the Alibi SD card is not supplied from Flintec, install a Flintec Alibi SD card.

Summary

Safety Instructions

Installation

Electrical Connections

Detailed wiring instructions for load cells and other interfaces.

Analogue Load Cell Connection (Only FT-111)

Wiring details for connecting analogue load cells to the FT-111.

Digital Load Cell Connection (Only FT-111 D)

Wiring details for connecting digital load cells to the FT-111 D.

Ethernet TCP;IP

Configuration and pinout for the Ethernet TCP/IP interface.

EtherNET;IP

Configuration and pinout for the EtherNET/IP interface.

FT-111(D) FUNCTIONS

Alibi Memory

Details on activating and using the Alibi memory for data recording.

PROGRAMMING AND CALIBRATION

Programming and Parameters

Navigating the programming menu and modifying parameters.

Calibration

Procedures for performing scale calibration with test weights.

DIGITAL LOAD CELL (DLC)

Addressing Digital Load cells

Steps for addressing and configuring digital load cells.

DIGITAL INPUTS AND OUTPUTS

SERIAL DATA OUTPUTS

OPTIONAL COMMUNICATION

Modbus RTU and TCP;IP

Details on Modbus RTU and TCP/IP communication protocols.

Ethernet TCP;IP

Configuration parameters for the Ethernet TCP/IP interface.

Profibus DP

Information on Profibus DP interface, baud rates, and LEDs.

Profinet

Details on the Profinet interface, LEDs, and configuration.

CANopen

Information on CANopen network communication, LEDs, and data format.

EtherCAT

Information on EtherCAT interface, LEDs, and data format.

CC-Link

Details on CC-Link network communication, LEDs, and configuration.

TROUBLE SHOOTING

Flintec FT-111 D Specifications

General IconGeneral
BrandFlintec
ModelFT-111 D
CategoryAccessories
LanguageEnglish