Questions and Answers

  • B
    Brian BradyAug 31, 2025
    What to do if Alibi memory SD card is not installed in Flintec FT-111 D?
    • J
      John SosaAug 31, 2025
      If the Alibi memory SD card is not installed, disable Alibi memory if it is not required. Otherwise, check the Alibi SD card or change the mainboard.
  • J
    Jason WattsAug 25, 2025
    What to do if the weight is too low for Flintec FT-111 D?
    • S
      shirleywarnerAug 25, 2025
      If the weight is too low, check the load, as the load cell or instrument could be broken.
  • J
    John BushAug 25, 2025
    What to do if there is a communication error between the display board and main board in Flintec Accessories?
    • C
      carlossmithAug 26, 2025
      If there is a communication error between the display board and the main board, power off the instrument and reenergize it after 30 seconds. If the issue persists, consider changing the main board.
  • B
    Brian CostaAug 26, 2025
    What to do if there is ADC initialization error in Flintec FT-111 D?
    • R
      Rebecca StoneAug 26, 2025
      If an ADC initialization error occurs, where the ADC could not initialize at power on due to a malfunction in the ADC or its interface circuitry, power off the instrument and reenergize it after 30 seconds. Also, check the external load cell connection and the load cell connector in the instrument. If problems persist, consider changing the main board or second scale board.
  • H
    Heidi SullivanAug 27, 2025
    What to do if the power source voltage is less than 9 VDC in Flintec Accessories?
    • J
      John HaleyAug 27, 2025
      If the power source voltage is less than 9 VDC, check the power supply voltage.
  • L
    Lisa LopezAug 27, 2025
    What to do if the load is not stable at calibration of Flintec FT-111 D Accessories?
    • D
      Daniel TorresAug 27, 2025
      If the load is not stable at calibration: * Wait until scale become stable. * Check grounding wiring. * Recalibrate the scale. * Change mainboard.
  • D
    Daniel RayAug 28, 2025
    What to do if NVM version error occurs in Flintec Accessories?
    • J
      James KingAug 28, 2025
      If an NVM version error occurs, causing factory defaults to load, press the Enter key if you changed the E2ROM (U13) to load the factory defaults. If you did not change it, consider changing the main board.
  • P
    phillipssusanAug 29, 2025
    What to do if there is Checksum error of calibration coefficients in Flintec Accessories?
    • K
      Kenneth MorenoAug 29, 2025
      If a checksum error of calibration coefficients occurs, check the scale performance. If necessary, recalibrate the scale or change the mainboard.
  • K
    Karen GonzalezAug 29, 2025
    What to do if there is Check sum error of parameters in Flintec FT-111 D?
    • E
      Elizabeth KnappAug 29, 2025
      If a checksum error of parameters occurs, check the scale performance. If necessary, recalibrate the scale or change the mainboard.
  • F
    feliciahuffAug 30, 2025
    What to do if the load cell output is decreased after loading in Flintec FT-111 D Accessories?
    • M
      Matthew ClineAug 30, 2025
      If the load cell output decreases after loading, check the load cell connections and the test weight loading. You may also need to recalibrate the scale or change the mainboard.

Summary

Safety Instructions

Installation

Electrical Connections

Detailed wiring instructions for load cells and other interfaces.

Analogue Load Cell Connection (Only FT-111)

Wiring details for connecting analogue load cells to the FT-111.

Digital Load Cell Connection (Only FT-111 D)

Wiring details for connecting digital load cells to the FT-111 D.

Ethernet TCP;IP

Configuration and pinout for the Ethernet TCP/IP interface.

EtherNET;IP

Configuration and pinout for the EtherNET/IP interface.

FT-111(D) FUNCTIONS

Alibi Memory

Details on activating and using the Alibi memory for data recording.

PROGRAMMING AND CALIBRATION

Programming and Parameters

Navigating the programming menu and modifying parameters.

Calibration

Procedures for performing scale calibration with test weights.

DIGITAL LOAD CELL (DLC)

Addressing Digital Load cells

Steps for addressing and configuring digital load cells.

DIGITAL INPUTS AND OUTPUTS

SERIAL DATA OUTPUTS

OPTIONAL COMMUNICATION

Modbus RTU and TCP;IP

Details on Modbus RTU and TCP/IP communication protocols.

Ethernet TCP;IP

Configuration parameters for the Ethernet TCP/IP interface.

Profibus DP

Information on Profibus DP interface, baud rates, and LEDs.

Profinet

Details on the Profinet interface, LEDs, and configuration.

CANopen

Information on CANopen network communication, LEDs, and data format.

EtherCAT

Information on EtherCAT interface, LEDs, and data format.

CC-Link

Details on CC-Link network communication, LEDs, and configuration.

TROUBLE SHOOTING

Flintec FT-111 D Specifications

General IconGeneral
BrandFlintec
ModelFT-111 D
CategoryAccessories
LanguageEnglish

Related product manuals