What to do if there is a Checksum error in Flintec FT-111 D?
- Llaura72Sep 1, 2025
If a checksum error occurs, check the termination resistors and the DLC connection (RS-485 & Power supply) hardware.
What to do if there is a Checksum error in Flintec FT-111 D?
If a checksum error occurs, check the termination resistors and the DLC connection (RS-485 & Power supply) hardware.
What to do if there is Checksum error of EPL printout format in Flintec FT-111 D Accessories?
If a checksum error of EPL printout format occurs, press enter key and re-load the EPL code, or change E2ROM ( U13 ).
What to do if the load cell output is decreased after loading in Flintec FT-111 D Accessories?
If the load cell output decreases after loading, check the load cell connections and the test weight loading. You may also need to recalibrate the scale or change the mainboard.
What to do if there is Checksum error of the data in Flintec FT-111 D?
If there is a checksum error of the data, change E2ROM ( U13 ).
What to do if there is E2PROM Error in Flintec Accessories?
If there is an E2PROM Error, change main board.
What to do if there is No regular response from load cell in Flintec FT-111 D Accessories?
If there is no regular response from the load cell, reenergize the indicator, check the DLC connection (RS-485 & Power supply), or change the load cell.
What to do if The DLC is over in Flintec Accessories?
If The DLC is over, check mechanical installation and DLC.
What to do if the power source voltage is less than 9 VDC in Flintec Accessories?
If the power source voltage is less than 9 VDC, check the power supply voltage.
What to do if there is ADC initialization error in Flintec FT-111 D?
If an ADC initialization error occurs, where the ADC could not initialize at power on due to a malfunction in the ADC or its interface circuitry, power off the instrument and reenergize it after 30 seconds. Also, check the external load cell connection and the load cell connector in the instrument. If problems persist, consider changing the main board or second scale board.
What to do if there is a communication error between the display board and main board in Flintec Accessories?
If there is a communication error between the display board and the main board, power off the instrument and reenergize it after 30 seconds. If the issue persists, consider changing the main board.
Detailed wiring instructions for load cells and other interfaces.
Wiring details for connecting analogue load cells to the FT-111.
Wiring details for connecting digital load cells to the FT-111 D.
Configuration and pinout for the Ethernet TCP/IP interface.
Configuration and pinout for the EtherNET/IP interface.
Details on activating and using the Alibi memory for data recording.
Navigating the programming menu and modifying parameters.
Procedures for performing scale calibration with test weights.
Steps for addressing and configuring digital load cells.
Details on Modbus RTU and TCP/IP communication protocols.
Configuration parameters for the Ethernet TCP/IP interface.
Information on Profibus DP interface, baud rates, and LEDs.
Details on the Profinet interface, LEDs, and configuration.
Information on CANopen network communication, LEDs, and data format.
Information on EtherCAT interface, LEDs, and data format.
Details on CC-Link network communication, LEDs, and configuration.