What to do if there is Check sum error of parameters in Flintec Accessories?
- KKayla SuarezAug 29, 2025
If a checksum error of parameters occurs, check the scale performance. If necessary, recalibrate the scale or change the mainboard.
What to do if there is Check sum error of parameters in Flintec Accessories?
If a checksum error of parameters occurs, check the scale performance. If necessary, recalibrate the scale or change the mainboard.
What to do if there is Checksum error of calibration coefficients in Flintec FT-111 Accessories?
If a checksum error of calibration coefficients occurs, check the scale performance. If necessary, recalibrate the scale or change the mainboard.
What to do if NVM version error occurs in Flintec Accessories?
If an NVM version error occurs, causing factory defaults to load, press the Enter key if you changed the E2ROM (U13) to load the factory defaults. If you did not change it, consider changing the main board.
What to do if the load is not stable at calibration of Flintec Accessories?
If the load is not stable at calibration: * Wait until scale become stable. * Check grounding wiring. * Recalibrate the scale. * Change mainboard.
What to do if the power source voltage is less than 9 VDC in Flintec FT-111 Accessories?
If the power source voltage is less than 9 VDC, check the power supply voltage.
What to do if there is ADC initialization error in Flintec FT-111 Accessories?
If an ADC initialization error occurs, where the ADC could not initialize at power on due to a malfunction in the ADC or its interface circuitry, power off the instrument and reenergize it after 30 seconds. Also, check the external load cell connection and the load cell connector in the instrument. If problems persist, consider changing the main board or second scale board.
What to do if there is a communication error between the display board and main board in Flintec Accessories?
If there is a communication error between the display board and the main board, power off the instrument and reenergize it after 30 seconds. If the issue persists, consider changing the main board.
What to do if the weight is too low for Flintec FT-111 Accessories?
If the weight is too low, check the load, as the load cell or instrument could be broken.
Brand | Flintec |
---|---|
Model | FT-111 Series |
Category | Accessories |
Language | English |
Comprehensive guide to various electrical connection options for the instrument.
Covers fulfilling data recording obligations using the alibi memory function.
Overview of the programming menu structure and parameter adjustment.
Detailed steps for performing scale calibration and adjustments.
Process for addressing digital load cells and shift adjustment methods.
Covers Modbus RTU and TCP/IP interfaces, data structures, and setup.
Details Ethernet TCP/IP interface parameters, data formats, and setup.
Information on Profibus DPV1 interface, baud rates, LEDs, and data format.
Details the Profinet interface, LEDs, parameters, and GSDML configuration.
Covers CANopen network communication, LEDs, and data format.
Details EtherNet/IP interface, LEDs, parameters, and EDS configuration.
Covers EtherCAT interface, LEDs, data format, and ESI configuration.
Details CC-Link network, LEDs, configuration, and data format.