FLIR CM275 USER MANUAL Document Identifier: CM275-en-US_AC
ACA function:
> 6A (rms) at 60.00A range for 10Hz ~ 1kHz
> 60A (rms) at 600.0A range for 10Hz ~ 1kHz
Hz function at ACA rotary switch:
> 6A (rms) for 10Hz ~ 10kHz
Table 13.4 Inrush current
Function Range Accuracy
ACA Inrush
60.00 A ±(3% + 0.3 A)
600.0 A ±(3% + 5 dgt)
Inrush current detection threshold: 0.5A for 60A range and 5.0A for 600.0A range
Overload protection: 1000 V
(rms)
, 600 A (rms).
Integration time is 100 ms.
Table 13.5 Resistance, Continuity, and Diode
Function Range Accuracy
Resistance
±(1.0% + 5 dgt)
Continuity 600.0 ±(1.0% + 5 dgt)
Diode 1.5V ±(1.5% + 5 dgt)
Overload protection: 1000 V
(rms)
.
Maximum test current: Approx. 0.1 mA.
: Approx. 1.8 V.
Maximum open circuit voltage for diode: Approx. 1.8 V.
Continuity threshold:
>
Continuity indicator: 2.7 kHz tone buzzer.
Continuity response time: <100 ms.