Advanced Thermal Solutions SC2500 User Manual
DCO65U-E SC2500 User Manual Page 19
How to define the set point value?
For best performances, we recommend you to choose a FPA Temperature
Set Point included in the following range: [Tcam – 20 °C; Tcam], Tcam
being the internal camera temperature as read at the bottom of ALTAIR’s
window.
If the camera’s internal temperature gets too high and makes your set
point to be out of the recommended range, you’ll have to manually
adjust the set point as described above.
Your SC2500 camera is delivered with several presets made with a
TEC set point of 35°C.