3 Fluke Corporation Fluke 1587/i400 Current Clamp FC Kit
https://www.fluke.com/th-th/product/electrical-testing/insulation-testers/fluke-1587-i400-fc-kit
600.0 V
Resolution 0.1 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(1% + 3)
Accuracy 60 Hz to 5000 Hz ±(% of
Rdg + Counts)
±(2% + 3)¹
1000 V
Resolution 1 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(2% + 3)
Accuracy 60 Hz to 5000 Hz ±(% of
Rdg + Counts)
±(2% + 3)¹
¹ 1 kHz bandwidth
Low-Pass Filter Voltage
Range
600.0 mV
Resolution 0.1 mV
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(1% + 3)
Accuracy 60 Hz to 400 Hz ±(% of Rdg
+ Counts)
+(2% + 3), -(6% - 3)
6.000 V
Resolution 0.001 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(1% + 3)
Accuracy 60 Hz to 400 Hz ±(% of Rdg
+ Counts)
+(2% + 3), -(6% - 3)
60.00 V
Resolution 0.01 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(1% + 3)
Accuracy 60 Hz to 400 Hz ±(% of Rdg
+ Counts)
+(2% + 3), -(6% - 3)
600.0 V
Resolution 0.1 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(1% + 3)
Accuracy 60 Hz to 400 Hz ±(% of Rdg
+ Counts)
+(2% + 3), -(6% - 3)
1000 V
Resolution 1 V
Accuracy 50 Hz to 60 Hz ±(% of Rdg
+ Counts)
±(2% + 3)
Accuracy 60 Hz to 400 Hz ±(% of Rdg
+ Counts)
+(2% + 3), -(6% - 3)
DC Voltage Measurement
Range Resolution Accuracy ±(% of Rdg + Counts)
6.000 V dc 0.001 V ±(0.09% + 2)