EasyManuals Logo

Fluke 16 User Manual

Fluke 16
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
16
Multimeter with Temperature
8
Table 3. Performance Tests
Test No.
Meter Range
Input To Meter
Switch
Position
Display
Minimum
Display
Maximum
1
2
3
4
5
6
7
8
9
1
10
4.000 V ac
4.000 V ac, MIN MAX
4.000 V ac
4.000 V ac
4.000 V ac
4.000 V ac
40.00 V ac
40.00 V ac
400.0 V ac
600 V ac
0 V
0 V
10 mV, 50 Hz
10 mV, 400 Hz
3.5 V, 50 Hz
3.5 V, 400 Hz
35 V, 50 Hz
35 V, 400 Hz
350 V, 80 Hz
600 V, 400 Hz
Center
Center
Center
Center
Right
Center
Center
Right
Right
Center
0.000
0.000
0.007
0.007
3.430
3.430
34.30
34.30
343.0
586
0.003
0.040
0.013
0.013
3.570
3.570
35.70
35.70
357.0
614
11
12
13
14
15
16
17
18
4.000 V dc, MIN MAX
4.000 V dc
40.00 V dc
400.0 V dc
4.000 V dc
40.00 V dc
400.0 V dc
600 V dc
0 V
0 V
+20 mV dc
-200 mV dc
+3.5 V
-35 V dc
+100 V dc
-600 V dc
Center
Center
Center
Center
Center
Center
Center
Center
-0.012
-0.002
00.01
-000.3
+3.466
-35.33
+99.0
-606
0.012
0.002
00.03
-000.1
+3.534
-34.67
+101.0
-594
19
20
21
22
23
24
25
26
27
200 µA dc
200 µA dc
200 µA dc
200 µA dc
200 µA dc
200 µA ac
200 µA ac
200 µA ac
200 µA ac
0.0 µA, dc
10.0 µA, dc
-10.0 µA, dc
200.0 µA, dc
-200.0 µA, dc
40.0 µA, 50 Hz
40.0 µA, 400 Hz
200.0 µA, 50 Hz
200.0 µA, 400 Hz
Right
Right
Right
Right
Right
Right
Right
Right
Right
-0.2
9.7
-10.3
197.8
-202.2
38.9
38.9
195.7
195.7
0.2
10.3
-9.7
202.2
-197.8
41.1
41.1
204.3
204.3
28
29
2
30
2
31
32
33
34
35
36
37
38
39
40
4
41
4
Continuity
Continuity, Automatic Selection
Continuity, Automatic Selection
400.0
400.0
400.0
4.000 k
40.00 k
400.0 k
4.000 M
40.00 M
, Automatic Selection
1.000 µF
1.000 µF
+2 V dc
2
+3.43 V dc
2
-0.5 V dc
2
0.0
1.0
350.0
1.0 k
35 k
100 k
2.9 M
35 M
1.96 V dc
0.0 µF
0.95 µF
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
Right
1.980
3.397
3
-0.507
3
0.0
0.8
346.6
0.990
34.67
99.0
2.873
34.44
1.940
3
-0.001 µF
0.930 µF
2.020
3.463
3
-0.493
3
0.2
1.2
353.4
1.010
35.33
101.0
2.927
35.56
1.980
3
0.001 µF
0.970 µF
1. In the Automatic Selection mode, the Meter uses a low-impedance thermistor (~2.5k for circuit protection and load testing
(referred to as low-Z input circuitry). When using the 5100B, 5500A, or 5700A to drive the Meter with high voltages, avoid an
overload/current limit condition by gradually stepping the voltage up (waiting two seconds between each step) from 90.0 V,
120.0 V, 180.0 V, and 350.0 V at 80 Hz each step.
2. Calibrator 50 divider override.
3. The dc volts annunciator must be on.
4. Conducting performance tests of the 400 , 4 k, 40 k, and 1 µF ranges (tests no. 22, 23, 24, 25, 26, 31, and 32) verifies
that the discrete and integrated circuitry needed to support the other capacitance ranges are working within specifications.
Therefore, the tests indirectly verify that the Meter will meet specification in the 10 µF, 100 µF, 1000 µF, and 10,000 µF
ranges.
Table 4. Temperature Performance Tests
Input Voltage Display Readings
°C °F
0.0 0.0 ± 0.8 32.0 ± 1.5
0.919 mV 23.0 ± 1.0 73.4 ± 2.2
4.095 mV 100.0 ± 1.8 212.0 ± 3.6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 16 and is the answer not in the manual?

Fluke 16 Specifications

General IconGeneral
TypeDigital Multimeter
Continuity CheckYes
Diode TestYes
Temperature Resolution0.1 °C
Accuracy±(0.5% + 1 digit)
Temperature Measurement-40 °C to 400 °C
Capacitance Measurement1000 µF
Battery Type9 V
Battery Life200 hours
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 60 °C
EmissivityNot applicable
Distance to Spot RatioNot applicable
AC Voltage Range0.1 mV to 600 V
DC Voltage Range0.1 mV to 600 V
Resistance Range0 - 40 MΩ

Related product manuals