Manual Supplement 1736/1738 Users
3 7/19
Change #3, 594
On page 16, replace the Measurement Line Power Source
section with:
Measurement Line Power Source:
XWWarning
To prevent injury, do not touch the metal parts of one test lead
when the other is still connected to hazardous voltage.
WCaution
To prevent damage to the Product, make sure the measured voltage
does not exceed the input rating of the power supply.
1. Attach the power supply to the Logger.
2. Move the slide-cover on the power supply to access the safety
sockets.
3. Connect the short test leads (see Figure 7B & 7C) with t
he
po
wer supply inputs. Make sure to use the non-stackable plugs.
The test leads are rated for measurement/overvoltage CAT III
1000 V and CAT IV 600 V.
4. Connect the test leads with the voltage measurement inputs:
Connect A/L1 with one input of the power supply.
Connect N with the second input of the power supply.
OR
Connect A/L1 with one input of the power supply.
Connect B/L2 with the second input of the power supply.
5. Use the short fan out of the Voltage Test Lead, 3-phase + N.
Plug the connector A/L1 into the socket A/L1 of the voltage
measurement inputs of the Logger. Repeat this with B/L2, C/L3
and N.