Performance Testing and Calibration
Performance Tests
4
4-5
Table 4-2. Volts Performance
Calibrator
Performance Limits
Output Fluke 39/41B V rms V pk V dc
V dc Range Min Max Min Max Min Max
5.00V 20 4.4 5.6
15.0V 50 14.4 15.6
65.0V 100 63.4 66.6
130.0V 200 127.1 132.9
208.0V 500 201 215
350.0V 500 340 360
480.0V 1000 467 493
600.0V 1000 585 615
V ac
5.00V 20 4.5 5.5 6.5 7.7
15.0V 50 14.7 15.3 20.5 22.0
65.0V 100 64.4 65.6 89.8 94.1
130.0V 200 129.1 130.9 179.9 187.8
208.0V 500 205 211 285 303
350.0V 500 346 354 483 507
480.0V 1000 475 485 662 695
600.0V 1000 595 605 828 868
4-7. Testing Amps Performance
Perform the following procedure to test the amps funtion of the Tester.
Warning
Ensure that the calibrator is in standby mode before making
any connection between the calibrator and Tester. dangerous
voltages may be preSent on the leads and connectors.
1. Press E and turn on the Tester to enter the Setup screen.
2. Using E and < > , select “OTHER” for Clamp Setting.
3. Press E to exit the Setup screen.
4. Connect a cable from the Output Voltage HI and LO connectors of calibrator “A” to
the Current Probe connector on the Tester.
5. Press M until the text screen is displayed on the Tester.
6. Press F until A is displayed in the upper right corner of the display.