Changing a Setup Option
1. Press or to scroll to the setup option you
want to change.
2. Press
to indicate that you want to change this
setting.
3. Press
or until the setting you want to use
appears on the display.
4. Press
to store the new setting in memory.
Notes
Setup is disabled in MIN MAX mode.
Offset:
The primary display shows the temperature plus
the offset and the secondary display shows the
offset. Remember to reset the offset to 0.0 when
it is no longer needed. The offset automatically
resets to 0.0 when you change the thermocouple
type. Model 52: You can store individual offsets
for T1 and T2.
Sleep mode:
The thermometer enters sleep mode if no button
press occurs for 20 minutes. Pressing any button
wakes the thermometer and returns it to its
previous state. Sleep mode becomes enabled
each time you turn on the thermometer and is
automatically disabled in MIN MAX mode.
Line frequency:
For optimum rejection of line noise, the
thermometer must be set for the local line
frequency.
1.888.610.7664 info@Fluke-Direct.com
Fluke-Direct
.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 51 II Series and is the answer not in the manual?

Fluke 51 II Series Specifications

General IconGeneral
Temperature Range-200°C to 1372°C (-328°F to 2501°F)
Resolution0.1°C or 0.1°F
Thermocouple CompatibilityJ, K, T, E
DisplayDigital, LCD
Battery Life1000 hours typical
Dimensions173 mm x 86 mm x 38 mm
Weight400 g
Operating Temperature0 °C to 50 °C
Storage Temperature-20 °C to 60 °C

Related product manuals