EasyManua.ls Logo

Fluke 712B - Ohms Source Specifications

Fluke 712B
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RTD Calibrator
Specifications
29
Ohms Source
Ohms Range
Excitation Current from
Measurement Device
Accuracy
(% of Output + Floor)
1 year 2 year
1.0 Ω to 400.0 Ω 0.1 mA to 0.5 mA 0.015 % + 0.1 Ω 0.03 % + 0.2 Ω
1.00 Ω to 400.00 Ω 0.5 mA to 3 mA 0.015 % + 0.05 Ω 0.03 % + 0.08 Ω
400.0 Ω to 1500.0 Ω 0.05 mA to 0.8 mA 0.015 % + 0.5 Ω 0.03 % + 0.8 Ω
1500.0 Ω to 4000.0 Ω 0.05 mA to 0.4 mA 0.015 % + 0.5 Ω 0.03 % + 0.8 Ω
Resolution
1.00 Ω to 400.00 Ω 0.01 Ω
400.0 Ω to 4000.0 Ω 0.1 Ω
Notes:
1) Supports pulsed transmitters and PLCs with pulse times as short as 5 ms.
2) Temperature coefficient: ± (0.002 % of output + 0.002 % of range) /°C ( <18 °C or >28 °C)