712B
Users Manual
30
RTD Input and Output
RTD Type
(α)
Range
(°C)
Measure (°C) Source (°C)
1 year 2 year
Source
Current
1 year 2 year
10Ω Pt(385)
-200 to 100 1.5 3 1 mA 1.5 3
100 to 800 1.8 3.6 1 mA 1.8 3.6
50Ω Pt(385)
-200 to 100 0.4 0.7 1 mA 0.4 0.7
100 to 800 0.5 0.8 1 mA 0.5 0.8
100 Ω
Pt(385)
-200 to 100
0.2 °C 0.4 °C
1 mA
0.2
°C
0.4 °C
100 to 800
0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
200 Ω
Pt(385)
-200 to 100
0.2 °C 0.4 °C
500 μA
0.2 °C 0.4 °C
100 to 630
0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
500 Ω
Pt(385)
-200 to 100
0.3 °C 0.6 °C
250 μA
0.3 °C 0.6 °C
100 to 630
0.015 %+0.28 °C 0.03 %+0.56 °C 0.015 %+0.28 °C 0.03 %+0.56 °C
1000 Ω
Pt(385)
-200 to 100
0.2 °C 0.4 °C
250 μA
0.2 °C 0.4 °C
100 to 630
0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C