Specifications
Specifications are based on a one year calibration cycle and
apply for ambient temperature from +18°C to +28°C unless
stated otherwise. “Counts” means number of increments or
decrements of the least significant digit.
Temperature Measure and Thermocouple Simulate
TC Type Resolution Error Reference
Junction
Error
J, K, T, E, L, U 0.1°C or °F ±(0.3°C + 10 µV) ±0.2°C
B, R, S 1°C or °F ±(0.3°C + 10 µV) ±0.2°C
Maximum input voltage: 30 V
Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.01 mV ±(0.025% + 1 count)
Maximum input voltage: 30 V