718Ex 30G/100G/300G
Users Manual
10
Table 2. Pushbutton Functions
Pushbutton Description
U
Press to select a different pressure unit. All units are available when the pressure sensor input is used. For
higher pressure module inputs, inappropriate (out-of-range) units are not available.
D
Turns pressure reading damping on and off. With damping on, the Calibrator averages several
measurements before displaying a reading.
Z
Press to zero the pressure display. Vent pressure to atmosphere before pressing this pushbutton. For an
Absolute Pressure Module, see Zeroing with Absolute Pressure Modules.
N
Press to read the minimum pressure and current readings since power was turned on or
c
was pressed.
Press again to read the maximum pressure and current readings since power was turned on or
c
was
pressed.
S Use for pressure switch test. See Switch Test.
c
Press to clear the MIN, MAX, and switch test memories
h
Press
h
to freeze the display. The
g
symbol appears on the display. Press
h
again to resume
normal operation.