718Ex 30G/100G/300G
Users Manual
22
Cleaning
Periodically wipe the case with a damp cloth; do not use
abrasives or solvents.
Calibration
Fluke recommends that the Calibrator be calibrated once
yearly to ensure that it performs according to its
specifications. A calibration manual is available.
Call 1-800-526-4731 from the U.S.A. and Canada. In other
countries, contact a Fluke Service Center.
Replacing the Battery
XW Warning
To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator
B
appears.
Remove the Calibrator from the Ex-hazardous
area before opening the battery door.
Use only the battery types listed in the
Approved Battery Table.
When
B
appears on the display, replace the 9 V
alkaline battery. Refer to Figure 7.
Hex Head
Wrench
alz008f.eps
Figure 7. Battery Replacement