718Ex 30G/100G/300G
Users Manual
26
Pressure Module Input
Range Resolution Accuracy
(determined by Pressure Module)
DC mA Input
Range Resolution
Accuracy, ±(% of
Reading + Counts)
24 mA 0.001 mA 0.02 + 2
Temperature coefficient: 0.005 % of range per
°
C for
temperature ranges -10
°
C to 18
°
C and 28
°
C to 55
°
C
General Specifications
Maximum voltage applied between either mA terminal and earth ground or between the mA terminals: 30 V
Pressure sensor media: Non-corrosive gasses only
Storage temperature: -40 °C to 71 °C
Operating temperature: -10 °C to 55 °C
Relative humidity: 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to 50 °C, and 35 % up to 55 °C
EMC: Complies with EN61326, Criteria C
Pollution Degree 2