Digital Multimeter
Specifications
5
Function Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )
AC Volts (Average
responding)
6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
2.0 % + 2
(45 Hz to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.3 % + 1
DC Volts 6.000 V
60.00 V
600.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
0.3 % + 1
Continuity 600 Ω 1 Ω Meter beeps at <25 Ω, beeper turns off at >250 Ω;
detects opens or shorts of 250 μs or longer.
Ohms
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
50.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.5 % + 2
0.5 % + 1
0.5 % + 1
0.5 % + 1
0.5 % + 1
2.0 % + 1
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
1000 nF
10.00 μF
100.0 μF
9999 μF
[1]
1 nF
0.01 μF
0.1 μF
1 μF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps (Average
responding)
[2]
60.00 mA
400.0 mA
[3]
6.000 A
10.00 A
[4]
0.01 mA
0.1 mA
0.001 A
0.01 A
2.5 % + 2
(45 Hz to 1 kHz)
Notes:
[1] In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
[2] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[3] 400.0 mA accuracy specified up to 600 mA overload.
[4] >10 A unspecified.
Function Range Resolution Accuracy ± ( [ % of Reading ] + [ Counts ] )
DC Amps
[3]
60.00 mA
400.0 mA
[4]
6.000 A
10.00 A
[5]
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 2
Hz
[1], [2]
(ac voltage input )
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 1
MIN MAX AVG For dc functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than
350 ms in duration.
For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than
350 ms in duration.
Notes:
[1] Frequency is specified from 2 Hz to 99.99 kHz.
[2] Below 2 Hz, the display shows zero Hz.
[3] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[4] 400.0 mA accuracy specified up to 600 mA overload.
[5] >10 A unspecified.