46
Table 14. Current Function Specifications (continued)
Accuracy
1
Function Range Resolution Model 83
2
Model 85
3, 4
Model 87
3, 4
Burden
Voltage
(typical)
µ
A
B
(45 Hz to 2 kHz)
µ
A
F
400.0 µA
4000 µA
400.0 µA
4000 µA
0.1 µA
1 µA
0.1 µA
1 µA
±(1.2% + 2)
5
±(1.2% + 2)
5
±(0.4% + 4)
±(0.4% + 2)
±(1.0% + 2)
5
±(1.0% + 2)
5
±(0.2% + 4)
±(0.2% + 2)
±(1.0% + 2)
±(1.0% + 2)
±(0.2% + 4)
±(0.2% + 2)
100 µV/µA
100 µV/µA
100 µV/µA
100 µV/µA
1. See the first sentence in Table 11 for a complete explanation of accuracy.
2. AC conversion for Model 83 is ac coupled and calibrated to the rms value of a sinewave input.
3. AC conversions for Models 85 and 87 are ac coupled, true rms responding, and valid from 5% to 100% of range.
4. See note 3 in Table 11.
5. Below a reading of 200 counts, add 10 counts.