Output Signal: 1 mV ac/A ac
Influence of Temperature on Accuracy: <0.15% per 10°C
(18°F) for temperatures from -10 to 18°C (14 to 64°F) and
from 28 to 50°C (82 to 122°F).
Basic Accuracy (45 to 65 Hz):
Input Current: Error: Phase Shift:
1 to 20A 5% of rdg + 0.3A Not Specified
20 to 100A 5% of rdg ±3°
100 to 500A 2% of rdg ±1°
Extended Accuracy:
For other frequencies, refer to the appropriate input current range
and add the error listed below to the “Basic Accuracy” error.
Input Current: Error:
(5 to 10 Hz): 1 to 500A not specified*
(10 to 20 Hz): 1 to 300A 5%
300 to 400A 15%
400 to 500A 25%
(20 to 45 Hz): 1 to 500A 5%
(65 Hz to 3 kHz): 1 to 50A 5%+0.4A
50 to 500A 5%
(3 kHz to 10 kHz): 1 to 500A not specified*
*-3 db @ 5 Hz and 10 kHz typical
Input Load Impedance (of host instrument): >1MΩ in parallel
with up to 47 pF
dV/dt max: 0.24 mV/µsec
Rise or Fall Time: <40 µsec
W Working Voltage: 600V ac rms, in compliance with
IEC-1010-1.
W Common Mode Voltage: 600V ac rms from earth ground, in
compliance with IEC 1010-1.
Influence of Adjacent Conductor: <5 mA/A
Influence of Conductor Position in Jaw Opening: +/-(1.5% of
reading +0.1A).
GENERAL SPECIFICATIONS
Dimensions: 66 x 195 x 34 mm (2.6 x 7.7 x 1.3 in.)
Weight: 420g (14.8 oz.)
Temperature: Operating: -10 to 50°C (14 to 122°F)
Storage: -40 to 71°C (-40 to 160°F)
Relative Humidity: 0 to 95% (10 to 30°C; 50 to 86°F)
0 to 75% (30 to 40°C; 86 to 104°F)
0 to 45% (40 to 50°C; 104 to 122°F)
Altitude: non-operating: 0 to 12000 meters (0
to 39370 ft.) operating: 0 to 2000
meters (0 to 6560 ft.)