83,85,87 Service Manual Supplement
2/01 1
Change #1
On pages 4-5, 4-11, and 4-17, add W to F1 and F2, add the following footnote to the bottom
of the table:
W To ensure safety, use exact replacement only.
On page 4-6, 4-12, and 4-18, add W near the callouts for F1 and F2.
Change #2
On page 3-11, Table 3-5, step 1, Display Reading:
From:
999.80 - 01000.2
To: 999.8 - 1000.2
On page 3-13, Table 3-9, Display Reading, change all three entries,
Step 2 From:
0.463 to 0.477
To: 0.462 to 0.478
Step 3 From: 0.0463 to 0.0477
To: 0.0462 to 0.0478
On page 3-14, Table 3-11, change AC Current for step 6,
From:
35.0 µA
To: 3500 µA