8505A
Table
2-4,
Error
Codes
MOMENTARY ERROR
CONDITIONS
(The reading in
progress is
aborted, but multimeter
operation is
automatically restored with the next
trigger. The
function annunciator
remains on during a
momentary error
condition.)
CODE
FAULT
SOLUTION
Error 0
V
DC/Ohm
Zero, zero attempted
in wrong function
(not
V DC
or OHMS) or an
overrange has been
entered.
Check
function. Only V DC and OHMS
are
permissible.
Error 1
Store attempted
during
overload
condition.
Change to
higher range
or (if storing
cal
correction
factors) use
lower
value source.
Error 6
Display
overflow error.
Check offset and
scaling
values.
Error 7
External Reference
error
-
voltage on one input
exceeds
20V dc.
Revise external
reference input.
Error 8
Controller
module is faulty.
Power off; replace
controller module.
Error
b
Illegal push button
sequence in
Calibration mode.
Wait till display
clears. Use
correct
sequence.
Error
C
invalid
push-button sequence,
or
illegal value
entered.
Wait till display
clears. Use correct
sequence
or value
within limits.
Error
d
Calibration
Memory
faulty or not
installed. Occurs
when
storing
into, or
recalling from,
Calibration
Memory,
or
at
power-up.
Replace
or install
Calibration
Memory
chip.
Error
F Cai
Memory
check-sum problem.
Try new power-up.
If necessary, reprogram Cal
Memory. Replacement
of Calibration Memory
may be required.
Error
H
Ohms
connection problem.
Verify that all connections
are proper. Check
input
fuses. Check input lead
polarity in
four-
terminal connections.
LATCHING ERROR
CONDITIONS
(A
valid function must be
selected to restore
multimeter operation
error condition.)
All function
annunciators
areoff during
a
latching
Error System error,
usually appears at power-up or
reset. Repeat
power-up or reset.
Error 2
Filter module faulty or not
installed.
Power off;
replace or
install filter module.
Error 3 DC
Signal
Conditioner module faulty or not
installed.
Power off; replace or
install DC Signal
Conditioner
module.
Error
4 OHMS,
A DC, or A AC
error.
Check for improper input
level. Check function
causing error
indication. Applicable
module
may need
replacement
or installation.
Error
5
Analog to Digital
Converter Module error.
Power off; replace
or install Analog to
Digital
Converter module.
Error 9
Function
selection error.
The function module
selected
is faulty
or not installed. V DC problem
causes
Error
3.
Select
valid function to clear
error condition.
Power off; replace or
install appropriate module.
Error E More
than one ac
converter, or the
wrong ac
converter, is
installed. Also
appears when
Calibration
Memory
module
is
installed.
Power off,
remove one ac converter,
or install
correct converter,
or remove the Calibration
Memory module
(the
multimeter
uses a
separate
Calibration Memory as
part of the Controller).
2-7