8500
Series
Table
2A-1. Programming
Instruction®
The programming instructions in this table pertain to the 8500 Series
Digital
Multimeters with the IEEE-488
Interface (Option -05), the
Bit Serial Interface
(Option -06)
or the Parallel Interface
(Option -07) installed.
Features and instructions unique to the DMM model or to the
Interface
used will be identified in the following
manner:
1.
8500A or 8502A: the symbol^ will denote an explanation
applicable
to
one DMM
model only. The
software version incorporated in the DMM may also be
mentioned for further identification. To verify
the
software version incorporated in your
instrument, observe the display indication at power on or reset. For
example, in the 8502A,
"HI-2.
0.2"
will appear in the display for models with software version 2.0.2.
2.
-05,
-06, or
-07
Interface Options: the symbol#
will
be used with a feature or
instruction unique
to a
particular Interface.
INITIALIZATION
When power is applied,
or the Reset character
(*)
is transmitted, the instrument assumes a preset default
condition. This condition is defined by the following remote codes:
REMOTE CODE
COMMAND
V
Volts DC
R4
1000V range
S5 (8500
A)
2
5
Samples per Reading
S7 (8502A)
2
7
Samples per Reading
F0
Fast
Filter, Timeout Disabled
(Panel Indicator OFF)
XO
External
Reference/Scaling
Disabled
P0
Offset
Feature Disabled
UO
Limits-Peak
Value Storage Disabled
TO
Single Reading Line
Synchronous
B0 Single Character ASCII Format
DO
Front
Panel Display Active
L0
Deactivate
Local Lockout
JO
Deactivate
Line Feed Suppression
M0
Enable
Cai Memory Factors
Q0
(8502A)
Disable External Trigger
W (8502A)
No Delay
#
Y0
(-06 Option only) Echo mode off (Bit Serial IF)
In
addition, the following instrument states are assumed
at power on
or Reset:
Remote/Local Local
Offset Zeroed
V dc Zero Zeroed or
*
Ohms
Zero Zeroed or
*
Cal Memory Factors
*
Peak Values
Cleared
Limits Values
Zeroed
Ext.
Ref/Scaling Values 1
$
8/16 Bit Mode
(-07 Option only)
8
Bit
‘Retained if Cal
Memory Option
-04
installed
2A-2