EasyManua.ls Logo

Fluke 8505A - General Specifications

Fluke 8505A
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8505A
Table
1-1. Options
OPTION
NO.
NAME
NOTES
01 AC/DC
Converter (Averaging)
1,
3
02A Ohms
Converter
03 Current Shunts
3
05 IEEE Standard 488-1975
Interface
2
06 Bit Serial Asynchronous
Interface
2
07 Parallel Interface
2
09A AC/DC Converter
(True RMS)
1,
3
1)
Options 01 and 09A
cannot
be
installed
simultaneously.
2)
Only
one of Options
05. 06.
and
07
can
be
installed
at any time.
3)
For the ac portion
of Option
03 to
operate,
either Option
01 or 09A
.....
must
be
installed.
Table
1-2.
Accessories
MODEL
OR
PART NO.
NAME
M04-205-600 574-inch Rack
Adapter
M00-260-610 18-inch Rack Slides
M00-280-61
0
24-inch Rack
Slides
80K-6 High Voltage
Probe
80K-40 High Voltage
Probe
83RF High Frequency
Probe
85RF High Frequency
Probe
Y8001 IEEE Std.
Cable, 1
Meter Length
Y8002 IEEE Std. Cable,
2 Meter
Length
Y8003 IEEE Std.
Cable, 4 Meter
Length
MIS-701 IK* Extender
Card
MIS-71 90K’ Static
Controller
MIS-7191K*
Test Module
MIS-7013K*
Bus Interconnect
and Monitor
Tor
use
during service
or repair.
j
Table
1-3.
Specifications
GENERAL
SPECIFICATIONS
Dimensions
10.8
cm High
x
43.2
cm Wide
x
42.5
cm Long
(4.25
in
High x 17 in Wide
x 16.75 in Long)
(See
Figure 1-1)
Weight
BASIC
... .
10 kg
(22
lbs)
FULLY LOADED
....
12
kg
(26
lbs)
Operating Power
VOLTAGE
. . . . 100V ac,
120V ac, 220V ac,
or 240V ac
(±10%)
BASIC INSTRUMENT POWER
. .
.
12 watts
FULLY LOADED POWER 24
watts
FREQUENCY
47 Hz
to 63 Hz
(400 Hz available on request)
Warm-Up
2 hours to rate
accuracy
Shock and
Vibration
Meets
requirements
of MIL-T-28800 for
type III, class
5,
style E
equipment.
Temperature Range
OPERATING
....
0°C
to
50°
C
NON-OPERATING ....
-40° C to
70°
C
Humidity Range
0°C TO 18°C . . . . 80% RH
18°C TO
40°
C . . . . 75% RH
40°
C
TO
50°
C . . . . 45% RH
Maximum
Terminal Voltage
LO TO GUARD
127V
rms
GUARD
TO CHASSIS
500V rms
HI SENSE TO HI SOURCE
....
127V rms
LO
SENSE TO LO SOURCE
.... 127V rms
HI SENSE TO LO SENSE
....
1000V rms
or 1200V dc
HI SOURCE TO LO SOURCE
.... 280V rms
1-2