8500
Series
Error
Codes (cont)
12 Filter Module Faulty or not
installed
13
DC
Signal Conditioner Module Faulty
or not installed
14
Excessive
voltage present
at terminals for
Ohms/Current Measurement
15
Fast A/D
Converter Faulty
or not installed
16
Numeric Display
Overflow
17 Magnitude of
External Reference Input
>20V
18 Controller Module Faulty
19
Function
Module selected not installed
Range Codes
The third character of the status response contains the following range information:
0
100
mV
dc,
100//A. 10O
1 IV dc, IV
ac,
1 mA, 1000
2
10V dc, 10V ac, 10 mA, Ikfl
3 100V dc, 100V
ac,
100 mA,
10 kO
4 1000V dc, 1000V ac, 1A, 100 kCl
5
1 MQ
6 10 MQ
7 100 MQ
Sample Codes
The fourth status response
character
contains sample information identified by the following codes:
0
1
2
3
4
5
6
7
1
Sample per Reading
2 Samples per Reading
4
Samples per Reading
8
Samples per Reading
16 Samples per Reading
32 Samples per
Reading
64 Samples per Reading
128 Samples per
Reading or
Greater
Function
Codes
5
<CR> <LF>
The fifth response character
identifies function:
0 DC
Volts
1 AC Volts
2 DC Amps
3 AC Amps
4 Ohms
5 DC Coupled AC Volts
7 Function Not Defined
2A-20