14 Fluke Corporation 8508A Extended Specifications
Read Rate and additional Uncertainty
Read Rate and Additional Uncertainty Specifications
ResolutionFunction
DCV, DCI &
Ohms
[10]
ACV & ACI
[6]
Normal Fast Normal Fast
Filter
Frequency
(Hz)
Read Rate
(readings/second)
Additional Errors
[13]
±
(ppm Reading +
ppm Range)
8 1/25 1/6 0 + 0 0 + 0.1
7 1/6 1/2 0 + 0.1 0 + 0.5
6 2 35 1.0 + 0.5 0 + 2.5
5 35 150 0 + 5 0 + 25
6 1 1/50 0 + 0
10 1/5 0 + 0
40 1/2 0 + 0
100 1 0 + 0
5 1 1/50 0 + 5
10 1/5 0 + 5
40 1/2 0 + 5
100 2 0 + 5
1 1/25 200 + 20
10 1/2.5 200 + 20
40 1 200 + 20
100 4 200 + 20
8 - 1/90 1/30 0 + 0
7 - 1/30 1/10 0 + 0.1
6 - 1/4 1/3 1.0 + 0.5
5 - 1/3 1/3 0 + 5