EasyManua.ls Logo

Fluke 8508A - AC Current Specifications

Fluke 8508A
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke Corporation 8508A Extended Specifications 9
AC Current specifications
AC Current
[1] [2] [6] [9]
Full Scale
Uncertainty Relative to Cal Stds
24 hour
TCal ±1 °C
365 day
TCal ±1 °C
Absolute Uncertainties
[9]
200 µA, 199.999 9 1 - 10 200 + 100 250 + 100 250 + 100 290 + 100 310 + 100
2 mA & 1.999 999
10 - 10k 200 + 100 250 + 100 250 + 100
280 + 100 300 + 100
20 mA 19.999 99 10k - 30k 500+ 100 600 + 100 600 + 100 650 + 100 710 + 100
30k - 100k 0.35 % + 100 0.4 % + 100 0.4 % + 100 0.4 % + 100 0.4 % + 100
200 mA 199.999 9
1 - 10 200 + 100 250 + 100 250 + 100 290 + 100 310 + 100
10 - 10k 200 + 100 250 + 100 250 + 100 250 + 100
290 + 100
10k - 30k 500 + 100 600 + 100 600 + 100 600 + 100 625 + 100
2 A 1.999 999
10 - 2k
500 + 100 600 + 100 600 + 100 600 + 100 620 + 100
2k - 10k 600 + 100 700 + 100 700 + 100 700 + 100 725 + 100
10k - 30k 0.25 % + 100 0.3 % + 100 0.3 % + 100 0.3 % + 100 0.3 % + 100
20 A
19.999 99
10 - 2k
700 + 100 800 + 100 800 + 100 800 + 100 820 + 100
2k - 10k 0.2 % + 100 0.25 % + 100 0.25 % + 100 0.25 % + 100 0.25 % + 100
200 µA, 199.999 9 1 - 10 250 + 120 300 + 120 300 + 120 380 + 120 400 + 120
2 mA & 1.999 999
10 - 10k 250 + 120 300 + 120 300 + 120
340 + 120 370 + 120
20 mA 19.999 99 10k - 30k 600+ 120 700 + 120 700 + 120 775 + 120 800 + 120
30k - 100k 0.35 % + 120 0.4 % + 120 0.4 % + 120 0.4 % + 120 0.4 % + 120
200 mA 199.999 9
1 - 10 250 + 120 300 + 120 300 + 120 380 + 120 400 + 120
10 - 10k 250 + 120 300 + 120 300 + 120 305 + 120
360 + 120
10k - 30k 600 + 120 700 + 120 700 + 120 700 + 120 740 + 120
2 A 1.999 999
10 - 2k
600 + 120 700 + 120 700 + 120 705 + 120 725 + 120
2k - 10k 700 + 120 800 + 120 800 + 120 815 + 120 860 + 120
10k - 30k 0.25 % + 120 0.3 % + 120 0.3 % + 120 0.3 % + 120 0.3 % + 120
20 A
19.999 99
10 - 2k
800 + 120 900 + 120 900 + 120 900 + 120 920 + 120
2k - 10k 0.2 % + 120 0.25 % + 120 0.25 % + 120 0.25 % + 120 0.25 % + 120
90 day
TCal ±1 °C
Range
95 % Confidence Level
99 % Confidence Level
Frequency
(Hz)
365 day
TCal ±1 °C
365 day
TCal ±5 °C
± (ppm Reading + ppm Range)
[4]