EasyManua.ls Logo

Fluke 8588A - Page 100

Fluke 8588A
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8588A/8558A
Operators Manual
96
Table 20. Memory Management Menu (cont.)
Menu Softkeys Parameter
(Save Record)
Push to select the current record into an archived record. Each push of this
softkey makes another archived record, as shown by the Stored Records field.
If the current record is still accumulating readings (in other words, the Product
is in the free-run mode), it continues to do so even after you push (Save
Record). Push (Manage Records) to view archived records.
(Manage Records)
Push to enter the Manage Records menu. The Manage Records menu shows
the archived records, which are stored as CSV files. See Figure 26. The
Records column shows the record file name and uses the date and time. The
latest record is shown on top. The # Readings column shows the number of
readings in each record. The Comment column contains user entered
comments using (Edit Comment) softkey. Comments are not stored in
the archived record, but only shows up in the Manage Records menu to help
identify records. The comment field is 15 characters wide. The Manage
Records softkeys are:
(Page Down): Used to view the archived records.
(Page Up): Used to view the archived records.
(Copy): Gives a submenu to copy a record into USB memory, with
these softkeys:
(Copy to USB): Copies the highlighted record into USB memory.
(Copy All to USB): Copies all of the archived records into USB
memory. The Product does not uniquely identify the USB ports. Insert only one
USB memory device during this operation. Push to move back to exit
this submenu.
(Edit Comment): Comments can be entered with the pop-up letter
keypad and , or the Product numeric keypad and . in the
numeric keypad is used for both numeric and letter entry.
iei047.png
Figure 26. Manage Records Menu

Table of Contents

Other manuals for Fluke 8588A

Related product manuals