 (Measure Setup): Aperture/PLC sets the integration time of the A to D 
converter using the navigational keys, much like Measure Setup in DCV. The 
choices are: 
•  Auto 
•  Auto Fast 
•  Manual 
When you select Manual, use the softkeys and the numerical keypad to edit the 
integration time by PLC and Time. The smallest time aperture is 0 seconds with 
200 ns increments and has an upper time limit of 10 seconds. The smallest 
aperture that can be set by PLC is 0.01. The upper limit is the PLC equivalent of 
10 seconds and is determined by the line frequency setting in the Instrument 
Setup menu. 
Shunt Corrections: When set to ON (Power-on default), the calculated current 
reading is based on the external shunt value and the shunt loading from the 
10 Mohm input impedance of the Product. Note that if set to OFF, an Instrument 
reset (Instrument Setup > Reset Instrument) retains the OFF setting. If the 
Product is power cycled, Shunt Corrections is always set to ON. 
Select Shunt Submenu 
The section explains the Select Shunt submenu.  
 (Page Down) and  (Page Up): Allows you to scroll through all of the 
current shunts stored in the Product. 
 (Sort By): Allows you to sort by Asset number, Serial number or Max A. 
Push  to cycle through the three choices. Note that the Basic shunt is 
always at the top. 
 (Delete Shunt): Allows you to delete the selected shunt (indicated by the 
darkened circle on the left). A user prompt shows before actual deletion.  
 (Manage Shunts): Allows you to edit specific characteristics of the shunt 
as well as adding a new shunt.  
Manage Shunts Submenu 
The Manage Shunts submenu is explained in this section. Enter the appropriate 
information for each of these fields with the navigation keys and numeric keypad.  
•  Asset number (shown as the first field on the shunt information line of 
the main DCI Ext Shunt screen) 
•  Manufacturer (displayed as the second field on the shunt information 
line) 
•  Model 
•  Serial Number 
•  Resistance value: Use the numeric keypad and  to input the 
resistance value of the shunt, for example from the most recent 
calibration certificate. Resistance value is shown as the fourth field in the 
shunt information line.