Reference Multimeter and 8 ½ Digit Multimeter 
  More 
67 
The Square, Triangle, and Truncated Sine selections are useful to measure the 
peak-to-peak output of multi-product calibrators like the Fluke 5522A which have 
these non-sine wave outputs. See the screen below: 
 
iei017.png 
The bottom field, Shunt Corrections ON/OFF determines if the ac-dc differences 
for the selected shunt are applied to the displayed current level, and the shunt 
loading is accounted for due to the input impedance of the voltage measurement 
circuit (10 Mohm in parallel with 80 pF). The main display indicates when 
corrections are ON. Note that if set to OFF, an Instrument reset (under 
Instrument Setup > Reset Instrument) retains the OFF setting.  If the Product 
is power cycled, Shunt Corrections is always set to ON.The Product uses a linear 
interpolation of the loaded ac-dc differences between the frequency points to 
make corrections. See the screen below: 
 
 
iei019.png